Detection of Visual Field Loss in Pituitary Disease: Peripheral Kinetic Versus Central Static



Rowe, Fiona J ORCID: 0000-0001-9210-9131, Cheyne, Christopher P, Garcia-Finana, Marta ORCID: 0000-0003-4939-0575, Noonan, Carmel P, Howard, Claire, Smith, Jayne and Adeoye, Joanne
(2015) Detection of Visual Field Loss in Pituitary Disease: Peripheral Kinetic Versus Central Static. NEURO-OPHTHALMOLOGY, 39 (3). pp. 116-124.

[img] Text
NO 2016 Uni deposit.pdf - Author Accepted Manuscript

Download (439kB)

Abstract

Visual field assessment is an important clinical evaluation for eye disease and neurological injury. We evaluated Octopus semi-automated kinetic peripheral perimetry (SKP) and Humphrey static automated central perimetry for detection of neurological visual field loss in patients with pituitary disease. We carried out a prospective cross-sectional diagnostic accuracy study comparing Humphrey central 30-2 SITA threshold programme with a screening protocol for SKP on Octopus perimetry. Humphrey 24-2 data were extracted from 30-2 results. Results were independently graded for presence/absence of field defect plus severity of defect. Fifty patients (100 eyes) were recruited (25 males and 25 females), with mean age of 52.4 years (SD = 15.7). Order of perimeter assessment (Humphrey/Octopus first) and order of eye tested (right/left first) were randomised. The 30-2 programme detected visual field loss in 85%, the 24-2 programme in 80%, and the Octopus combined kinetic/static strategy in 100% of eyes. Peripheral visual field loss was missed by central threshold assessment. Qualitative comparison of type of visual field defect demonstrated a match between Humphrey and Octopus results in 58%, with a match for severity of defect in 50%. Tests duration was 9.34 minutes (SD = 2.02) for Humphrey 30-2 versus 10.79 minutes (SD = 4.06) for Octopus perimetry. Octopus semi-automated kinetic perimetry was found to be superior to central static testing for detection of pituitary disease-related visual field loss. Where reliant on Humphrey central static perimetry, the 30-2 programme is recommended over the 24-2 programme. Where kinetic perimetry is available, this is preferable to central static programmes for increased detection of peripheral visual field loss.

Item Type: Article
Uncontrolled Keywords: Humphrey, kinetic, octopus, perimetry, pituitary, static
Depositing User: Symplectic Admin
Date Deposited: 15 Jul 2016 13:34
Last Modified: 19 Jan 2023 07:33
DOI: 10.3109/01658107.2014.990985