Kamel, Hanan, Syms, Richard RA, Kardoulaki, Evdokia M and Rea, Marc ORCID: 0000-0001-9982-1703
(2018)
Surgical wound monitoring by MRI with a metamaterial-based implanted local coil.
EPJ APPLIED METAMATERIALS, 5.
p. 5.
Abstract
<jats:p>An implantable sensor for monitoring surgical wounds after bowel reconstruction is proposed. The sensor consists of a coupled pair of 8-element magneto-inductive ring resonators, designed for mounting on a biofragmentable anastomosis ring to give a local increase in signal-to-noise ratio near an annular wound during <jats:sup>1</jats:sup>H magnetic resonance imaging. Operation on an anti-symmetric spatial mode is used to avoid coupling to the <jats:italic>B</jats:italic><jats:sub>1</jats:sub> field during excitation, and a single wired connection is used for MRI signal output. The electrical response and field-of-view are estimated theoretically. Prototypes are constructed from flexible elements designed for operation at 1.5 T, electrical responses are characterized and local SNR enhancement is confirmed using agar gel phantoms.</jats:p>