SERS-based lateral flow immunoassay utilising plasmonic nanoparticle clusters for ultra-sensitive detection of salivary cortisol



Oh, Hyun-Kyung, Akbali, Baris, Sham, Tung-Ting ORCID: 0000-0002-9648-4964, Haworth-Duff, Adam, Blair, Joanne C, Smith, Barry L, Sricharoen, Nontawat, Lima, Cassio ORCID: 0000-0002-9062-6298, Chen, Tsan-Yao, Huang, Chen-Han
et al (show 4 more authors) (2025) SERS-based lateral flow immunoassay utilising plasmonic nanoparticle clusters for ultra-sensitive detection of salivary cortisol NANOSCALE, 17 (34). pp. 19656-19665. ISSN 2040-3364, 2040-3372

[thumbnail of D5NR02062J.pdf] Text
D5NR02062J.pdf - Author Accepted Manuscript
Available under License Creative Commons Attribution.

Download (1MB) | Preview

Abstract

Cortisol plays a central role in maintaining physiological homeostasis, and both cortisol excess and deficiency are associated with life-threatening conditions. Accurate diagnosis, adequate treatment and monitoring of disorders of cortisol secretion are essential for good health, normal growth and development. Although commercially available lateral flow immunoassay (LFI) strips can be used to measure cortisol, they have limitations, especially low sensitivity and limited quantitative performance, inhibiting their use in clinical settings. Here, we present a novel LFI platform integrated with surface-enhanced Raman scattering (SERS), employing precisely size-controlled gold nanoparticle clusters functionalised with Raman reporter molecules to overcome these limitations. The approach achieves exceptional sensitivity, covering the relevant therapeutic range in humans, with a limit of detection (LOD) for cortisol of 0.014 pg mL−1, which is >500 times more sensitive than conventional LFI strips. The platform also showed high specificity for cortisol. The diagnostic potential was confirmed by testing with human saliva samples (n = 28), cross-validated with UPLC-MS/MS, showing excellent correlation (R2 = 0.9977). Bland-Altman analysis demonstrated strong agreement, with all samples falling within the 95% limits and yielding a mean bias of −3.5% ± 13.2% relative to UPLC-MS/MS. Given its sensitivity, specificity and simplicity, this LFI-SERS platform offers strong potential for clinical translation to enable convenient cortisol monitoring.

Item Type: Article
Uncontrolled Keywords: Saliva, Humans, Gold, Hydrocortisone, Immunoassay, Spectrum Analysis, Raman, Metal Nanoparticles, Limit of Detection
Divisions: Faculty of Science & Engineering
Faculty of Science & Engineering > School of Electrical Engineering, Electronics and Computer Science
Depositing User: Symplectic Admin
Date Deposited: 31 Jul 2025 10:03
Last Modified: 16 Jun 2026 17:07
DOI: 10.1039/d5nr02062j
Related Websites:
URI: https://livrepository.liverpool.ac.uk/id/eprint/3193829
Disclaimer: The University of Liverpool is not responsible for content contained on other websites from links within repository metadata. Please contact us if you notice anything that appears incorrect or inappropriate.