Lawman, S, Madden, PW, Romano, V ORCID: 0000-0002-5148-7643, Dong, Y, Mason, S, Williams, B ORCID: 0000-0001-5930-287X, Kaye, S ORCID: 0000-0003-0390-0592, Willoughby, C, Harding, S ORCID: 0000-0003-4676-1158, Shen, Y ORCID: 0000-0002-8915-1993 et al (show 1 more authors)
(2017)
Deformation velocity imaging using optical coherence tomography and its applications to the cornea.
Biomedical Optics Express, 8 (12).
pp. 5579-5593.
This is the latest version of this item.
Text
AsFinalSubmission.pdf - Author Accepted Manuscript Available under License : See the attached licence file. Download (2MB) |
|
Text
RomanoBiomedOptExp2017.pdf - Published version Available under License : See the attached licence file. Download (4MB) |
Abstract
Optical coherence tomography (OCT) can monitor human donor corneas non-invasively during the de-swelling process following storage for corneal transplantation, but currently only resultant thickness as a function of time is extracted. To visualize and quantify the mechanism of de-swelling, we present a method exploiting the nanometer sensitivity of the Fourier phase in OCT data to image deformation velocities. The technique was demonstrated by non-invasively showing during de-swelling that osmotic flow through an intact epithelium is negligible and removing the endothelium approximately doubled the initial flow at that interface. The increased functional data further enabled the validation of a mathematical model of the cornea. Included is an efficient method of measuring high temporal resolution (1 minute demonstrated) corneal thickness, using automated collection and semi-automated graph search segmentation. These methods expand OCT capabilities to measure volume change processes for tissues and materials.
Item Type: | Article |
---|---|
Uncontrolled Keywords: | (110.4500) Optical coherence tomography, (120.5050) Phase measurement, (170.4500) Optical coherence tomography, (170.6935) Tissue characterization |
Depositing User: | Symplectic Admin |
Date Deposited: | 13 May 2020 10:04 |
Last Modified: | 18 Jan 2023 23:52 |
DOI: | 10.1364/BOE.8.005579 |
Related URLs: | |
URI: | https://livrepository.liverpool.ac.uk/id/eprint/3086533 |
Available Versions of this Item
-
Deformation velocity imaging using optical coherence tomography and its applications to the cornea. (deposited 11 Oct 2017 06:49)
- Deformation velocity imaging using optical coherence tomography and its applications to the cornea. (deposited 13 May 2020 10:04) [Currently Displayed]